M.53 MA Aqueous Shunts and Stents for Glaucoma
MA Aqueous Shunts and Stents for Glaucoma
M.53
M.53 MA Aqueous Shunts and Stents for Glaucoma
Description
Glaucoma is characterized by elevated intraocular pressure (IOP), which results in visual field loss and irreversible blindness if left untreated. In the primary (conventional) outflow pathway from the eye, aqueous humor passes through the trabecular meshwork, enters a space lined with endothelial cells (Schlemm canal), drains into collector channels, and then into the aqueous veins. Increases in resistance in the trabecular meshwork and/or the inner wall of the Schlemm canal can disrupt the balance of aqueous humor inflow and outflow, resulting in an increase in IOP and glaucoma risk.
Surgical procedures for glaucoma aim to reduce intraocular pressure (IOP) resulting from impaired aqueous humor drainage in the trabecular meshwork and/or Schlemm canal. In the primary (conventional) outflow pathway from the eye, aqueous humor passes through the trabecular meshwork, enters a space lined with endothelial cells (Schlemm canal), drains into collector channels, and then into the aqueous veins. Increases in resistance in the trabecular meshwork and/or the inner wall of the Schlemm canal can disrupt the balance of aqueous humor inflow and outflow, resulting in an increase in IOP and glaucoma risk.
Dates
-
Original Effective
05-01-2026
-
Last Review
11-11-2026
-
Next Review
11-04-2027
Policy
I. Insertion of ab externo aqueous shunts approved by the U.S. Food and Drug Administration may be considered reasonable and necessary as a method to reduce intraocular pressure in individuals with glaucoma where medical therapy has failed to adequately control intraocular pressure.
II. Use of an ab externo aqueous shunt for all other conditions, including individuals with glaucoma when intraocular pressure is adequately controlled by medications, is considered not reasonable and necessary because the effectiveness of this procedure has not been established.
III. Insertion of ab interno aqueous stents approved by the U.S. Food and Drug Administration as a method to reduce intraocular pressure in individuals with glaucoma where medical therapy has failed to adequately control intraocular pressure may be considered reasonable and necessary.
IV. Implantation of 1 or 2 U.S. Food and Drug Administration-approved ab interno stents in conjunction with cataract surgery may be considered reasonable and necessary in individuals with mild-to-moderate open-angle glaucoma treated with ocular hypotensive medication.
V. Use of ab interno stents for all other conditions is considered not reasonable and necessary because the effectiveness of this procedure has not been established
Background
Surgical intervention may be indicated in patients with glaucoma when the target IOP cannot be reached pharmacologically. Trabeculectomy (guarded filtration surgery) is the most established surgical procedure for glaucoma, which involves dissecting the conjunctiva, creating a scleral flap and scleral ostomy then suturing down the flap and closing the conjunctiva, allowing aqueous humor to directly enter the subconjunctival space. This procedure creates a subconjunctival reservoir, which can effectively reduce IOP, but commonly results in filtering “blebs” on the eye, and is associated with numerous complications (eg, hemorrhage, scarring, hypotony, infection, leaks, bleb-related endophthalmitis) and long-term failure. Other surgical procedures (not addressed herein) include trabecular laser ablation, deep sclerectomy (which removes the outer wall of the Schlemm canal and excises deep sclera and peripheral cornea), and viscocanalostomy (which unroofs and dilates the Schlemm canal without penetrating the trabecular meshwork or anterior chamber). Canaloplasty involves dilation and tension of the Schlemm canal with a suture loop between the inner wall of the canal and the trabecular meshwork. This ab externo procedure uses the iTrack illuminated microcatheter (iScience Interventional) to access and dilate the entire length of the Schlemm canal and to pass the suture loop through the canal.
Currently, minimally invasive glaucoma surgeries (MIGS) are alternative, less invasive techniques that are being developed and evaluated. Similar to trabeculectomy, the objective of MIGS is to lower IOP by improving outflow of eye fluid; however, MIGS involves less surgical manipulation of the sclera and the conjunctiva compared than a trabeculectomy.
MIGS can either be performed outside the eye (ab externo) or inside the eye (ab interno). Examples of ab externo devices cleared by the U.S. Food and Drug Administration (FDA) include the Ahmed, Baerveldt, Molteno, and EX-PRESS mini-shunt, which shunt aqueous humor between the anterior chamber and the suprachoroidal space. These devices differ by explant surface areas, shape, plate thickness, presence or absence of a valve, and details of surgical installation. Generally, the risk of hypotony (low pressure) is reduced with aqueous shunts compared with trabeculectomy, but IOP outcomes are worse than after standard guarded filtration surgery. Complications of anterior chamber shunts include corneal endothelial failure and erosion of the overlying conjunctiva. The risk of postoperative infection is lower with shunts than with trabeculectomy, and failure rates are similar (≈10% of devices fail annually). The primary indication for aqueous shunts is for failed medical or surgical therapy, although some ophthalmologists have advocated their use as a primary surgical intervention, particularly for selected conditions such as congenital glaucoma, trauma, chemical burn, or pemphigoid.
Examples of ab interno devices either approved or given marketing clearance by FDA include the iStent, which is a 1-mm long stent inserted into the end of the Schlemm canal through the cornea and anterior chamber; the CyPass suprachoroidal stent; and XEN gelatin stent. Because aqueous humor outflow is pressure-dependent, the pressure in the reservoir and venous system is critical for reaching the target IOP. Therefore, some devices may be unable to reduce IOP below the pressure of the distal outflow system used.
Clinical Rationale
Surgical procedures for glaucoma aim to reduce intraocular pressure (IOP) resulting from impaired aqueous humor drainage in the trabecular meshwork and/or Schlemm canal. In the primary(conventional) outflow pathway from the eye, aqueous humor passes through the trabecular meshwork, enters a space lined with endothelial cells (Schlemm canal), drains into collector channels, and then into the aqueous veins. Increases in resistance in the trabecular meshwork and/or the inner wall of Schlemm canal can disrupt the balance of aqueous humor inflow and outflow, resulting in an increase in IOP and glaucoma risk.
More recently the Trabectome™, an electrocautery device with irrigation and aspiration, has been used to selectively ablate the trabecular meshwork and inner wall of the Schlemm canal without external access or creation of a subconjunctival bleb. IOP with this ab interno procedure is typically higher than the pressure achieved with standard filtering trabeculectomy. Canaloplasty involves dilation and tension of the Schlemm canal with a suture loop between the inner wall of the canal and the trabecular meshwork. This ab externo procedure uses the iTrack™ illuminated microcatheter (iScience Interventional) to access and dilate the entire length of the Schlemm canal and to pass the suture loop through the canal.
Aqueous shunts may also be placed in the anterior or posterior chamber to facilitate drainage of aqueous humor. Established shunts include theAhmed™ (New World Medical), Baerveldt® (Advanced Medical Optics), Molteno® (IOP), ExPress® mini-shunt (Alco), and the SOLX® DeepLight® Gold Micro-Shunt (SOLX), which shunts aqueous humor between the anterior chamber and the suprachoroidal space. These devices differ depending on explant surface areas, shape, plate thickness, the presence or absence of a valve, and details of surgical installation. Generally, the risk of hypotony (low pressure) is reduced with aqueous shunts in comparison with trabeculectomy, butIOP outcomes are higher than after standard guarded filtration surgery. Complications of anterior chamber shunts include corneal endothelial failure and erosion of the overlying conjunctiva. The risk of postoperative infection is less than after trabeculectomy, and failure rates are similar, with about 10% of devices failing each year. The primary indication for aqueous shunts is when prior medical or surgical therapy has failed, although some ophthalmologists have advocated their use as a primary surgical intervention, particularly for selected conditions such as congenital glaucoma, trauma, chemical burn, or pemphigoid.
The Xen gel stent(from AqueSys) is a soft, permanent, non-migrating, subconjunctival implant that shunts fluid from the anterior chamber to the subconjunctival space. It is a compressible and tissue-conforming hydrophilic implant made of porcine gelatin that's crosslinked with gluteraldehyde, a material that's been used for other medical purposes. The cylindrical, 6-mm long implant comes with three differerent-size lumens: 45 um ; 63 um ; and 140 um.
The iStent Inject from Glaukos-one of two new variations on the current iStent-is pushed directly through the trabecular meshwork into Schlemm's canal.
The Hydrus Microstent from Ivantis is a novel, 8-mm long Schlemm's canal scaffold made out of nitinol, a biocompatible alloy. It has two principal mechanisms: providing a bypass of the trabecular meshwork and scaffolding Schlemm's canal for three clock hours where it can access multiple collector channels. To implant the Hydrus, you approach the trabecular meshwork and Schlemm's canal under a gonio lens; with the inserter you puncture the trabecular meshwork and slide the device into the canal. Thus, part of the device is in the anterior chamber, while the majority of the device is in the Schlemm's canal. This maintains a bypass of the trabecular meshwork similar to the iStent.
The CyPass Micro-Stent (Transcent Medical). This is another ab interno glaucoma device. It's inserted through the angle into the supraciliary space, designed to take advantage of the negative pressure gradient between the suprachoroidal space and the anterior chamber. It's made from a polyimide material; it's 6.35 mm long, with an external diameter of 510 um.
Practice Guidelines and Position Statements
American Glaucoma Society
In 2020, the American Glaucoma Society published a position paper on microinvasive glaucoma surgery. The Society supports efforts that facilitate patient access to these procedures, including more flexible regulatory pathways for new devices, expansion of the indications for already approved devices, and greater availability of information obtained by regulatory authorities.
National Institute for Health and Care Excellence
The National Institute for Health and Care Excellence (2017) updated guidance on trabecular stent bypass microsurgery for open-angle glaucoma (OAG). The guidance stated that “Current evidence on trabecular stent bypass microsurgery for OAG raises no major safety concerns. Evidence of efficacy is adequate in quality and quantity."
The National Institute for Health and Care Excellence (2018) published guidance entitled "Microinvasive subconjunctival insertion of a trans-scleral gelatin stent for POAG". The guidance states that evidence is limited in quantity and quality and therefore, the procedure should only be used with special arrangements and that patients should be informed of the uncertainty of the procedure
Quick Code Search
Procedure
Diagnosis
Codes
Insertion of anterior segment aqueous drainage device, without extraocular reservoir, internal approach, into the suprachoroidal space
INSERTION OF AQUEOUS DRAINAGE DEVICE, WITHOUT EXTRAOCULAR RESERVOIR, INTERNAL APPROACH, INTO THE SUBCONJUNCTIVAL SPACE; INITIAL DEVICE
INSERTION OF AQUEOUS DRAINAGE DEVICE, WITHOUT EXTRAOCULAR RESERVOIR, INTERNAL APPROACH, INTO THE SUBCONJUNCTIVAL SPACE; EACH ADDITIONAL DEVICE (LIST SEPARATELY IN ADDITION TO CODE FOR PRIMARY PROCEDURE)
INSERTION OF ANTERIOR SEGMENT AQUEOUS DRAINAGE DEVICE, WITH CREATION OF INTRAOCULAR RESERVOIR, INTERNAL APPROACH, INTO THE SUPRACILIARY SPACE
Insertion of anterior segment aqueous drainage device into the trabecular meshwork, without external reservoir, and without concomitant cataract removal, one or more
AQUEOUS SHUNT TO EXTRAOCULAR EQUATORIAL PLATE RESERVOIR, EXTERNAL APPROACH; WITHOUT GRAFT
AQUEOUS SHUNT TO EXTRAOCULAR EQUATORIAL PLATE RESERVOIR, EXTERNAL APPROACH; WITH GRAFT
Insertion of anterior segment aqueous drainage device, without extraocular reservoir, external approach
REVISION OF AQUEOUS SHUNT TO EXTRAOCULAR EQUATORIAL PLATE RESERVOIR; WITHOUT GRAFT
Revision of aqueous shunt to extraocular equatorial plate reservoir; with graft
Extracapsular cataract removal with insertion of intraocular lens prosthesis (1 stage procedure), manual or mechanical technique (eg, irrigation and aspiration or phacoemulsification); with insertion of intraocular (eg, trabecular meshwork, supraciliary, suprachoroidal) anterior segment aqueous drainage device, without extraocular reservoir, internal approach, one or more
Ocular implant, aqueous drainage assist device
Aqueous shunt
References
|
2024
Panarelli JF, Moster MR, Garcia-Feijoo J, et al. Ab-Externo MicroShunt versus Trabeculectomy in Primary Open-Angle Glaucoma: Two-Year Results from a Randomized, Multicenter Study. Ophthalmology. Mar 2024; 131(3): 266-276. PMID 37769852 |
|
2024
Swaminathan SS, Jammal AA, Medeiros FA, et al. Visual Field Outcomes in the Primary Tube Versus Trabeculectomy Study. Ophthalmology. Oct 2024; 131(10): 1157-1163. PMID 38582154 |
|
2024
Matsuo M, Fukuda H, Buathong J, et al. Comparison of 1-year effectiveness between phaco-microhook ab-interno trabeculotomy and phaco-iStent trabecular micro-bypass stent in primary open-angle glaucoma with low-teen intraocular pressure. Graefes Arch Clin Exp Ophthalmol. Aug 19 2024. PMID 39160440 |
|
2023
Park J, Rittiphairoj T, Wang X, et al. Device-modified trabeculectomy for glaucoma. Cochrane Database Syst Rev. Mar 13 2023; 3(3): CD010472. PMID 36912740 |
|
2023
Tokumo K, Okada N, Onoe H, et al. Ex-PRESS Implantation versus Trabeculectomy for Long-Term Maintenance in Patients with Open-Angle Glaucoma. Clin Ophthalmol. 2023; 17: 2525-2537. PMID 37662650 |
|
2023
Sheybani A, Vera V, Grover DS, et al. Gel Stent Versus Trabeculectomy: The Randomized, Multicenter, Gold-Standard Pathway Study (GPS) of Effectiveness and Safety at 12 Months. Am J Ophthalmol. Aug 2023; 252: 306-325. PMID 36972738 |
|
2022
Lim SY, Betzler BK, Yip LWL, et al. Standalone XEN45 Gel Stent implantation in the treatment of open-angle glaucoma: A systematic review and meta-analysis. Surv Ophthalmol. 2022; 67(4): 1048-1061. PMID 35081414 |
|
2021
Konopinska J, Byszewska A, Saeed E, et al. Phacotrabeculectomy versus Phaco with Implantation of the Ex-PRESS Device: Surgical and Refractive Outcomes-A Randomized Controlled Trial. J Clin Med. Jan 22 2021; 10(3). PMID 33499300 |
|
2018
Chaudhary, A, Salinas, L, Guidotti, J, et al. XEN Gel Implant: a new surgical approach in glaucoma. Expert Review of Medical Devices. 2018;15(1):47-59. |
|
2017
Hengerer FH, Kohnen T, Mueller M, et al. Ab interno gel implant for the treatment of glaucoma patients with or without prior glaucoma surgery: 1-year results. J Glaucoma. Dec 2017;26(12):1130-1136. PMID 29035911 |
|
2017
De Gregorio A, Pedrotti E, Russo L, et al. Minimally invasive combined glaucoma and cataract surgery: clinical results of the smallest ab interno gel stent. Int Ophthalmol. May 29 2017. PMID 28555256 |
|
2017
Grover, DS, Flynn, WJ, Bashford, KP, et al. Performance and Safety of a New Ab Interno Gelatin Stent in Refractory Glaucoma at 12 Months. Am J Ophthalmol 2017 Nov;183:25-36. |
|
2015
Pfeiffer N, Garcia-Feijoo J, Martinez-de-la-Casa JM, et al. A Randomized Trial of a Schlemm's Canal Microstent with Phacoemulsification for Reducing Intraocular Pressure in Open-Angle Glaucoma. Ophthalmology. Jul 2015;122(7):1283-1293. PMID 25972254 |
|
2014
Netland PA, Sarkisian SR, Jr., Moster MR, et al. Randomized, prospective, comparative trial of EX-PRESS glaucoma filtration device versus trabeculectomy (XVT study). Am J Ophthalmol. Feb 2014;157(2):433-440 e433. PMID 24210765 |
|
2014
Fea AM, Belda JI, Rekas M, et al. Prospective unmasked randomized evaluation of the iStent inject ((R)) versus two ocular hypotensive agents in patients with primary open-angle glaucoma. Clin Ophthalmol. 2014;8:875-882. PMID 24855336 |
|
2013
Hoeh H, Ahmed, II, Grisanti S, et al. Early postoperative safety and surgical outcomes after implantation of a suprachoroidal micro-stent for the treatment of open-angle glaucoma concomitant with cataract surgery. J Cataract Refract Surg. Mar 2013;39(3):431-437. PMID 23506920 |
|
2012
Gedde SJ, Schiffman JC, Feuer WJ, et al. Treatment outcomes in the Tube Versus Trabeculectomy (TVT) study after five years of follow-up. Am J Ophthalmol. May 2012;153(5):789-803 e782. PMID 22245458 |
|
2012
Craven ER, Katz LJ, Wells JM, et al. Cataract surgery with trabecular micro-bypass stent implantation in patients with mild-to-moderate open-angle glaucoma and cataract: Two-year follow-up. J Cataract Refract Surg. Aug 2012;38(8):1339-1345. PMID 22814041 |
|
2012
Belovay GW, Naqi A, Chan BJ, et al. Using multiple trabecular micro-bypass stents in cataract patients to treat open-angle glaucoma. J Cataract Refract Surg. Nov 2012;38(11):1911-1917. PMID 22980724 |
|
2012
Boland MV, Ervin AM, Friedman D, et al. Treatment for Glaucoma: Comparative Effectiveness. Comparative Effectiveness Review No. 60. (Prepared by the Johns Hopkins University Evidence-based Practice Center under Contract No. HHSA 290-2007-10061-I.) AHRQ Publication No. 12-EHC038-EF. Rockville, MD: Agency for Healthcare Research and Quality; 2012 |
|
2011
de Jong L, Lafuma A, Aguade AS, et al. Five-year extension of a clinical trial comparing the EX-PRESS glaucoma filtration device and trabeculectomy in primary open-angle glaucoma. Clin Ophthalmol. 2011;5:527-533. PMID 21607021 |
|
2011
Francis BA, Singh K, Lin SC, et al. Novel glaucoma procedures: a report by the American Academy of Ophthalmology. Ophthalmology. Jul 2011;118(7):1466-1480. PMID 21724045 |
|
2011
Samuelson TW, Katz LJ, Wells JM, et al. Randomized evaluation of the trabecular micro-bypass stent with phacoemulsification in patients with glaucoma and cataract. Ophthalmology. Mar 2011;118(3):459-467. PMID 20828829 |
|
2020
Fellman RL, Mattox C, Singh K, et al. American Glaucoma Society Position Paper: Microinvasive Glaucoma Surgery. Ophthalmol Glaucoma. 2020; 3(1): 1-6. PMID 32672638 |
|
2017
National Institute for Health and Care Evidence (NICE). Trabecular stent bypass microsurgery for open-angle glaucoma [IPG575]. 2017; https://www.nice.org.uk/guidance/ipg575. |
|
2018
National Institute for Health and Care Excellence. Microinvasive subconjunctival insertion of a trans-sceral gelatin stent for primary open-angle glaucoma. [IPG612]. 2018; https://www.nice.org.uk/guidance/ipg612/chapter/1-Recommendations. |
Revisions
|
11-17-2026
Adding cpt 66991 to policy |
|
11-11-2026
annual review at MPC- no changes |